Eliminate flow rate pauses and drop-offs common when push-pull style pumping systems change directions. With this "Next Generation" system, the refilling syringe fills at a faster rate then the infusing syringe, giving it time to prime the syringe and/or overlap the infusion. This creates a near seamless transition between syringes. The infusing pump changing direction, or reaching overlap volume, triggers the paused pump to start infusing.
Eliminates Flow Rate Pauses & Drop-OffsThe DUAL-NE-1000X Continuous Pump System eliminates the problem of flow rate drop-offs and pauses found in "Push-Pull" syringe pump systems. With "Push-Pull" systems, one syringe infuses, while the other refills, then they switch directions. This switch of directions causes a delay in flow while the syringes are primed.
The DUAL-NE1000X Continuous Pump System solves the inconsistent flow rate problems of "push-pull" pumps.
How it works:
The refilling syringe pump fills the syringe at a faster rate than the infusing syringe pump infuses, giving it time to prime the syringe after the syringe is filled, or even overlap the infusion to create a seamless transition between syringes.
The smoothness of the transition between syringes can be fine tuned by adjusting the infusion rate, refill rate, priming rate, prime volume and overlap volume parameters.
| Syringe Size | Maximum Rate | Minimum Rate |
| 0.5 µL | 25.49 µL/hr | 0.001 µL/hr |
| 1 mL | 52.86 mL/hr | 0.727 µL/hr |
| 3 mL | 223.8 mL/hr | 3.076 µL/hr |
| 5 mL | 372.5 mL/hr | 5.119 µL/hr |
| 10 mL | 607.6 mL/hr | 8.349 µL/hr |
| 20 mL | 966.2 mL/hr | 13.28 µL/hr |
| 30 mL | 1260 mL/hr | 17.32 µL/hr |
| 60 mL | 2120 mL/hr | 29.1 µL/hr |
Special Application: Continuous Infusion (Push-Pull) and Dual Syringe Pump Using 2 NE-4000 pumps attached with a cable, one pump will infuse while the other pump refills or have simultaneous infusion.
| Model | Description | Price |
| FW-1-X | X Firmware for stand-alone pumps (NE-1000 series) | $130 USD |
| FW-5-X | X Firmware for OEM style pumps (NE-500 series) | $75 USD |
| FW-1-X2 | X2 Firmware for stand-alone pumps (NE-1000 series) | $230 USD |
| FW-5-X2 | X2 Firmware for OEM style pumps (NE-500 series) | $130 USD |
| CBL-DUAL-3 | Cable to connect 2 pumps to use the Dual Pumps Modes | $10 USD |
| FW-EXTRACTOR | PLCC chip extractor tool for replacing firmware | $4 USD |