CombiFlash Torrent and Torrent Aq

Versatile Large-Scale Flash Purification CombiFlash Torrent® systems from Teledyne ISCO provides the ease of use and high reliability of our CombiFlash and RediSep® product lines in a large, development-scale platform. Packed with time-saving features and controlled by our innovative PeakTrak® software, the CombiFlash Torrent accommodates a broad range of injection, detection, and collection requirements. Choose from two models: CombiFlash Torrent for a combination of normal phase and reverse phase purifications, or CombiFlash Torrent AQ optimized for reverse phase.
Patented active solvent sensing technology monitors solvent supply and waste levels for true walk-away automation. Pressure and vapor sensors ensure safe operation by stopping the purification when pressure increases, or a leak is detected. Fully grounded tubing allows safe operation at high flow rates. Optional air purging of the column after use simplifies disposal.

Safety

Patented active solvent sensing technology monitors solvent supply and waste levels for true walk-away automation. Pressure and vapor sensors ensure safe operation by stopping the purification when pressure increases, or a leak is detected. Fully grounded tubing allows safe operation at high flow rates. Optional air purging of the column after use simplifies disposal.


Versatility

Tailor a system to meet your needs with a variety of sample introduction and fraction handling configurations.

Highly Programmable and Easy to Use

Teledyne ISCO’s powerful yet intuitive PeakTrak software—the industry standard in flash chromatography—lets you quickly access and edit every method parameter from one screen.

Standard Features

  • - 1 L/min flow rate
  • - 100 psi rated
  • - 700 g purifications
  • - Patented solvent and waste level sensing technology
  • - Fully-grounded fluid path
  • - Easy-to-use PeakTrak® software
  • - Accepts RediSep column sizes from 80 g to 7 g silica and 100 g to 8.6 kg C18
  • - RFID column detection
  • - Automated six-port valve and large diameter tubing

Specifications

Flow Rate Range 5 - 200 mL/min (Flash & Prep)
Pump Prime Self-priming
Maximum Pressure (Flash operation) 200 psi (13.8 bar)
Maximum Pressure (Prep operation) 3,500 psi (240 bar)
UV and UV-Vis Detection UV 200-400 nm (standard); UV-Vis 200-800 nm (optional)
ELSD Detection Built-in ELSD detector compatible with UV or UV-Vis
Mass Spectrometer Three models with mass ranges of 10-1,200 and 10-2,000 Da
Programmable gradients Linear, step, isocratic
Sample introduction (Flash) Solid or liquid load
Sample introduction (Prep) Variable volume liquid load (5mL loop standard)