Optizen Alpha

OPTIZEN Alpha is a spectrophotometer that uses a double-beam method.OPTIZEN Alpha can measure the transmittance or absorbance at each wavelength of a sample in ultraviolet and visible light bands to determine the quantitative characteristics such as concentration and purity.

OPTIZEN Alpha, which can be used widely from general analytical experiments to professional research fields, guarantees accurate measurement and excellent reproducibility and provides reliable results in various fields such as the environment, biotechnology, and chemistry.

The previous single beam type spectrophotometer has a weak point of error occurrence in measuring a sample, because of the movement of a light source by time lag in measuring the strength of between a reference light and a light from sample. To solve the problem, OPTIZEN Alpha was designed as a double-beam type spectrophotometer. The system utilizes an additional reference beam to improve the measurement performance by compensating the intensity fluctuation of its light source.


Main Charateristics

  • A world-class measurement performance
  • Wide-size color screen (ALPHA: 8”)
  • Various cell compatibility and fast cell type choice
  • Automatic measurement of lots of samples by equipping multi-cell.
  • Convenient voice service and volume control
  • Emotional design

Product Features

  • User Convenience
    All functions of existing PC software are installed in OPTIZEN Alpha, making it faster and more convenient.
  • Incredible Speed
    Fast and flexible software delivers the best result with analytical speeds up to 1.5 times that of existing products.
  • Rapid Service
    Quick services are available as OPTIZEN Alpha is made in Korea from design conception to manufacturing based on the unique technology of K LAB Co., Ltd.
  • Extensive ScalabilityOPTIZEN Alpha is designed for high-precision/high-resolution measurement in widebands from ultraviolet to visible light and can be used in various applications.
  • Compact SizeCompact OPTIZEN Alpha enhances the efficiency of the experimental space.
  • Sensuous DesignNow, you can enjoy a greater level of pleasure with a sensuous design.

Specifications

.
Photometrics System Double-beam type
Monochromator Czerny-Turner type with 1200 lines/nm blazed grating
Light Source Tungsten Halogen Lamp & Deuterium Lamp
Lamp interchange Wavelength 340~410 nm (Default 370 nm)
Detector Silicon Photodiodes
Spectral Bandwidth 1 nm (190 to 1100 nm)
Wavelength Range 190 to 1100 nm
Wavelength Setting 0.05 nm
Wavelength Accuracy ±0.3 nm (For entire range)
±0.1 nm (656.1 nm)
Wavelength Repeatability <± 0.1 nm
Slew Rate About 13,500 nm/min
Scan Speed max 6,000 nm/min
Photometric Range Absorbance : -4 to 4 Abs
Transmittance : 0 % to 400 %
Photometric Accuracy ±0.002 Abs at 0.5 Abs
±0.004 Abs at 1.0 Abs
±0.006Abs at 2.0Abs
Photometric repeatability ± 0.0002 at 0.5 Abs
± 0.0006 at 1.0 Abs
± 0.001 at 2.0 Abs
Baseline stability < 0.0003 Abs/h
Baseline flatness <± 0.0005 Abs
Stray light < 1.0 % at 198 nm (KCl)
< 0.02 % at 220 nm (NaI)
< 0.02 % at 340 nm (NaNO₂)
Standard cell Holder Autometic Rotary type 8-position Multi-Cell Holder
Operating system(OS) Windows 10 (Embedded PC);
Display 8-inch color LCD with touch screen
Power Requirement 100~240 V; 50~60 Hz
Dimensions(WxDxH) 520(W)x500(D)x200(H)
Weight(kg) 14 kg
PC Software (optional) OPTIZEN View for Windows®
(optional) OPTIZEN Secure™